Indus OA 303 Oxygen Analyser consists of a heated zirconia sensor and a technology.It gives a linear(Nemstain) output for the entire range from sub-ppm to percentage levels of oxygen.

 

Principle
Principle High temp. zirconia sensor operating at about at about 750°C is used to sense the oxygen level. The sensor gives mV output, which is converted to %O2 & displayed



Technical  Data
Heated Sensor
Sensing Element
  Stabilised Zirconia fused to alumina tube
Thermocouple
  K type (for measuring sensor temp)
Heater
 
Built-in heater
Power Supply   Mains power
Power
 
20 W + Heater(250W)
Dimensions & Weight
Analyser
 
345 mm x 305 mm x 130 mm
Weight
  
2 kgs
Heated Probe   90 x 90 x 380mm,3kg
Transformer Case   120 x 160 x 200 mm,4kg


Image for oms103


Technical  Data
Analyser
Type   Digital,based on microcontroller technology
Measurement
 
Oxygen,Temperature
Data Range        
 
0 to 21% Oxygen
Temperature
 
30 to 1000°C
Resolution
 
0.1% Oxygen,1°C Temperature
Repeatability
 
Better than 0.2%.
Keyboard
 
Membrane keypad with 12 keys
Programming
 
Password protection,set points,calibration.
Inputs   Oxyprobe(mV) & Thermocouple connections
Output    
Standard
  16 characters x 2 line backlit LCD display
Optional
  RSC232-C interface
Process Temperature   Upto 1200°C continuous operation,upto 1400°C discontinuous operation
Start up time   Within 30 seconds of power ON
Response time   < 250 ms.
Alarms   LED Display and audible alarm
Set points   Two alarm set points, which can be set to any value by keypad,with relay outputs

< Next >