Indus
Model BAA-601Breath
Analyser
The INDUS BAA 601 breath alcohol analyzer incorporates the latest technology
for accurate measurement of breath alcohol content.
With an Infra-Red (wireless) printing option, the equipment is most
suited for use in mobile patrolling vehicles. A range of 30 metres gives
you the flexibility to print a report even if you are standing at a
distance.
The printer unit is connected to the vehicle battery to eliminate the
requirement of regular power sources. The equipment itself functions
on rechargeable batteries to reduce consumable cost
|
Technical
Data
|
|
How
to use
|
| 1) |
|
Switch
on the device. |
| 2) |
|
Wait
for 10 sec and then Press ZERO key to perform zero setting |
| 3) |
|
The unit is now ready for operation. Ask the person being tested to
blow into the mouthpiece for at least 2 seconds. |
| 4) |
|
M
Press SAMPLE to accept reading |
| 5) |
|
Repeat
this to obtain 3 readings |
| 6) |
|
Press the key PRN to send the data to the remote printer for printing |
| 7) |
|
The printer will print the values, average and the statutory limits |
| 8) |
|
Wait till the reading comes to zero before testing again. |
The instrument has an in built low
battery detection & warning system. When the Low Batt. LED glows red,
batteries have to be changed or recharged.
|
|
Technical
Data
|
| Measuring
Unit |
| Operating
Voltage |
|
9
V |
| Range
of Measurement |
|
0.00
- 0.10%BAC |
|
Warm Up Time |
|
<
60 sec |
| Response
Time |
|
2
Sec |
| Sensor |
|
Semiconductor
type alcohol sensor |
| Remote
Printer |
| Operating
Voltage |
|
12
V |
| Impact
Printer |
|
24
Columns |
| RF
Frequency |
|
433.9
MHz (Licence Free) |
| Range
of Operation |
|
30
meters |
| Standard
Accessories |
| |
|
Disposable
mouth piece |
| |
Sampling
tube (6" length) |
| Optional
Accessories |
|
|
Rechargeable
batteries |
| |
Battery
charger. |
| Range |
|
Audio
& Visual Indication |
| 0.00%
~ 0.01% |
|
1
beep sound |
| 0.01%
~ 0.03% |
|
1
beep sound |
| 0.03%
~ 0.05% |
|
2
beep sounds together with yellow LED light |
| 0.05%
~ 0.10% |
|
Beep
sound for 5 sec together with red LED |
| Key
Functions |
| ZERO |
|
Zero
Setting |
|
SAMPLE |
|
Samples
can be taken by pressing this key. The average of the samples will
decide the end result. |
|
PRN |
|
To
print the samples at the Remote printer. |
|
1 |
|
Used
for calibration of the equipment |
|